3000 Micro GC Operating Manual

Poor sensitivity (continued) Impure carrier gas

Ensure the carrier gas purity is between
99.995% and 99.9995%.
Add hydrocarbon/moisture/oxygen traps or
particulate filters to the carrier gas line to
improve purity.

Internal leak

Contact

INFICON

.

Malfunctioning column or
injector

Check the flow out of the analytical and
reference vents on the back of
3000 Micro GC. If there is no flow from either
or both vents, contact

INFICON.

Malfunctioning TCD

Contact

INFICON

.

Flat line chromatogram

Inject Time set to 0 ms.

In EZ IQ, change the Inject Time to a value
other than 0 ms.

Sample Pump turned off

In EZ IQ, ensure that the Sample Pump
checkbox is selected and set for at least 15
seconds.

Malfunctioning column or
injector

Check the flow coming out of the
3000 Micro GC analytical and reference
vents. If there is no flow from either or both
vents, contact

INFICON.

Malfunctioning sample pump Listen for the noise from the pump when

taking a sample. If there is no noise, contact

INFICON

.

Malfunctioning TCD

Contact

INFICON

.

Missing peak

Backflush Time set too
early (if applicable)

Increase the Backflush Time until all peaks of
interest appear on the chromatogram.

Saturated peak

Inject Time too long

In EZ IQ, shorten the Inject Time.

Injector Temperature too
high

In EZ IQ, lower the Injector Temperature.

Detector Sensitivity set to
High

In EZ IQ, set the Detector Sensitivity to
Standard.

Table 9-1 Common chromatographic problems (continued)

Symptom

Possible Cause

Remedy