5 - 13

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

5.1.5.3 Installation Instructions for the Pressure Reducer and Genie Filter Assembly

1

Turn off 3000 Micro GC and wait for the sample inlet fitting to cool.

WARNING

3000 Micro GC has inlets which can be heated to 140°C.
Touching the inlets at operating temperatures can result
in injury. Extreme care should be taken to avoid touching
the heated inlets.

2

Move the 3000 Micro GC handle fully away from the front cover.

3

Remove any fittings or hardware currently installed on the 3000 Micro GC
sample fitting.

4

Place a 1/4 in. male nut, back ferrule, and front ferrule onto one end of the outlet
tube provided. See

Figure 5-16

for ferrule orientation. See

Figure 5-17

for outlet

tube orientation.

5

Loosely install a 5/16 in. nut, back ferrule, and front ferrule onto the sample inlet
fitting. Do not tighten. The outlet tube will be installed into this fitting later.

Figure 5-16 Installing ferrules

1/4 in. male nut

Back ferrule

Front ferrule

Back ferrule

Front ferrule

5/16 in. female nut