8.3.2 Running a Low Pressure, Clean Gas Sample

To obtain a low pressure, clean gas sample, use the standard 10 micron sample
inlet filter assembly for filtration (PN G2801-60900, included in the ship kit).

1

Turn 3000 Micro GC on.

2

Make sure the carrier gas supply is sufficient to run all of the samples.

3

If using a sample vessel, prepare the sample.

NOTE: If not installed, install the sample inlet filter assembly. Refer to

section

4.2.5.3 on page 4-13

.

4

Connect the sample container or gas source to the 3000 Micro GC sample
inlet.

NOTE: For sample source pressures above 172 kPa (25 PSI), install a pressure

regulator as shown in

Figure 8-29

.

Figure 8-29 Example sample connection for a clean, low pressure sample

5

Open the sample vessel valve, and adjust the pressure to <172 kPa (<25 PSI).

6

Use EZ IQ software to load the desired method and perform the analysis.

7

When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.

8

Repeat steps 3 to 7 to run the next sample.

Set input pressure
0 to 172 kPa
(0 to 25 PSI)

Recommended
35 to 70 kPa
(5 to 10 PSI)

Safety stand
for cylinder

Dual-end ferrule

External filter assembly

G2801-60900