8.2.6 Install a GC Module

Once a module is removed, it can be replaced with another module. If the
replacement module uses a different carrier gas, follow the procedure in

section

6.4, Change Carrier Gas Configuration, on page 6-9

.

NOTE: For the Standard GC Module, skip step 1 and proceed to step 2.

1

Set the Performance Enhanced GC module address by setting the DIP switch
on the upper right of the D-board to the appropriate configuration. See

Figure

8-17

.

Figure 8-17 DIP switch position on the D-board

Each setting determines a unique module address. Refer to the label on the top
of the module. DEF is the default DIP switch setting that is set in the factory.
See

Figure 8-18

.

Figure 8-18 Module address label on top of a performance enhanced module