8.3.4 Running a High Pressure Gas Sample without Liquids/Particulates

1

If the Pressure Reducer (PN G2815A) is not installed, install one as instructed
in

section 5.1.4, Installing the Pressure Reducer, on page 5-8

. See

Figure 8-31

.

NOTE: The Pressure Reducer can connect directly to the 3000 Micro GC or to

the sample inlet filter (PN G2801-60900).

Figure 8-31 Pressure reducer sample conditioner (PN G2815A)

2

Turn 3000 Micro GC on.

3

Check that the carrier gas supply is sufficient to run all of the samples.

4

If using a sample vessel, prepare the sample.

5

Connect the sample conditioner or gas source to the sample input fitting on the
Pressure Reducer accessory.

6

Open the sample vessel valve.

7

Use EZ IQ software to load the desired method and perform the analysis.

8

When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.

9

Repeat steps 4 to 8 to run the next sample.

Sample in

To sample input fitting
on 3000 Micro GC

Vent

Adjustment knob

Mounting bracket not shown