4.2.6 Power and Communication Cable Setup

1

Connect the crossover cable (PN 5183-4649) between the computer and
3000 Micro GC.

NOTE: If a LAN connection through a hub is required, consult an IT specialist

to configure the setup.

2

Connect the power cable to the AC/DC converter supplied in the Shipping Kit.

3

If desired, connect the remote start cable (PN G2801-60618) to
3000 Micro GC. See

section 5.2, The Remote Connector, on page 5-30

for

remote start cable pinout details.

4

If a customized checkout is necessary or no checkout is required, reconfigure
carrier gas type as needed by following the instructions in

section 6.4, Change

Carrier Gas Configuration, on page 6-9

.

Figure 4-9 Power and communication connections

Crossover cable

(PN 5183-4649)

Standalone Installation

Power cord and converter

Remote start/stop
cable

(PN G2801-60618)

Crossover cable

(PN 5183-4649)