6 - 15

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

7

A window will prompt to restart 3000 Micro GC. Click Restart and wait at least
three minutes before shutting down the 3000 Micro GC. See

Figure 6-21

.

CAUTION

Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.

Figure 6-21 IP configuration - restart

8

Turn off 3000 Micro GC.

9

If needed, reconfigure the computer’s IP address for local LAN use.

10

If needed, disconnect the crossover cable, and connect the 3000 Micro GC and
the computer to the local LAN using standard LAN cables (PN G1530-61495).

11

If needed, reboot the computer.

12

Turn on 3000 Micro GC.

13

Reconnect to 3000 Micro GC using the new IP address. Ping the new IP
Address. For example, if the new IP address is 10.1.1.102, click Start >> Run
and type ping 10.1.1.102. Click OK to ping the instrument. See

Figure 6-22

.

Figure 6-22 Ping new IP address