1 rga checkout gas – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 267

Advertising
background image

11 - 19

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

11.5.1 RGA Checkout Gas

Refinery Gas Analyzer (RGA) Checkout Gas may be used during installation of a
3000 Micro GC configured for refinery gas composition analysis to verify chemical
performance.

Table 11-11

lists the component concentrations.

NOTE: Do not use RGA Checkout Gas as a calibration standard for specific

refinery gas applications. Since refinery gas composition varies greatly
among various petrochemical processes, acquire a calibration standard
that meets specific requirements.

Table 11-11 RGA checkout gas

Components

Concentration (Mole %)

hydrogen

12.0%

oxygen

0.2%

nitrogen

balance

carbon monoxide

1.0%

carbon dioxide

3.0%

methane

5.0%

ethane

4.0%

ethylene

2.0%

acetylene

1.0%

propane

2.0%

propylene

1.0%

1,2-propadiene

1.0%

methyl acetylene

1.00%

isobutane

0.3%

n-butane

0.3%

1-butene

0.3%

isobutylene

0.3%

trans-2-butene

0.3%

cis-2-butene

0.3%

1,3 butadiene

0.3%

isopentane

0.1%

n-pentane

0.1%

1-pentene

0.1%

cis-2-pentene

0.1%

trans-2-pentene

0.1%

2-methyl-2-butene

0.05%

n-hexane

0.05%

Advertising