3000 Micro GC Operating Manual

TCD detector 'fault'
(continued)

Malfunctioning column or
injector

Check the flow coming out of the
3000 Micro GC analytical and reference
vents. If there is no flow from either or both
vents, contact

INFICON

.

Malfunctioning TCD

Contact

INFICON

.

Method fields do not appear Instrument not properly

configurated

Re-configure the instrument through EZ IQ
Config
.

Conflicting module
addresses

Ensure the DIP switch is set to the correct
address for enhanced modules.

Corrupted software file

Re-install EZ IQ. Refer to the
074-537 EZ IQ Installation Guide.

Table 9-2 Common software problems (continued)

Symptom

Cause

Remedy