Module. see, Figure 8-6, Caution – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual
Page 191
Advertising
8 - 9
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
Figure 8-6 Decommission a GC module - remove options
7
Click Shutdown on the confirmation message. See
Figure 8-7 Decommission a module - shut down
CAUTION
Do not turn off the power without waiting at least three
minutes. 3000 Micro GC must update configuration files
before power is turned off. If power is turned off too soon,
the files could be corrupted, making 3000 Micro GC
inoperable.
8
Wait at least three minutes before shutting down 3000 Micro GC. See
.