INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 199

Advertising
background image

8 - 17

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

2

Remove the small metal plate covering the O-rings on the upper gang block.
See

Figure 8-19

.

Figure 8-19 Protective plate on upper gang block

3

Inspect the GC module mounting flange fitting to verify all new O-rings on the
replacement GC module are undamaged and seated properly. See

Figure

8-20

.

Figure 8-20 Inspect the new O-rings

Retaining screw

Metal plate covering O-rings and
upper gang block

Mounting flange

Verify O-ring placement

Gang block, upper half

Advertising