7.15 Apply the Valco Valve Program to EZ IQ Method

7.15.1 Valco Valve Setup in EZ IQ

1

In EZ IQ, click Method >> Instrument Setup.

2

Click the Sampling tab. See

Figure 7-68

.

Figure 7-68 Instrument setup window - switching

3

Select the Valve purge position: from the drop-down list. This is the initial
valve position for purging before starting a single run in a sequence.

4

Type a Purge delay time:. This is the time defined for switching the valve to a
position which is specified in the Valve purge position after starting a run.

5

Save the method by clicking File >> Method >> Save.