2.3.5 Repeatability

Typical results for Relative Standard Deviation (RSD) at constant temperature and
pressure (for C

1

to C

6

components at % level) are shown below for peak areas.

Injector Type

Peak Area Repeatability

Variable volume . . . . . . . . . . . . . . . . <1% RSD

Fixed volume . . . . . . . . . . . . . . . . . . <0.2% RSD

Backflush (timed mode) . . . . . . . . . . <1% RSD

Backflush (fixed mode). . . . . . . . . . . <0.5% RSD

2.3.6 Carrier Gas Specifications

Connection to 3000 Micro GC . . . . . 1/8 in. Swagelok fittings

External source . . . . . . . . . . . . . . . . Helium

Hydrogen
Nitrogen
Argon

Carrier gas purity . . . . . . . . . . . . . . . 99.995 to 99.9995%

Input pressure . . . . . . . . . . . . . . . . . 552 to 566 kPa (80 to 82 PSI)

2.3.7 Portable 3000 Micro GC Carrier Gas Specifications

Onboard cylinder source . . . . . . . . . Helium

Nitrogen
Argon

Onboard cylinder capacity . . . . . . . . 12,410 kPa (1800 PSI)

Onboard cylinder usage . . . . . . . . . . 35-40 hours, application dependent

2.3.8 External Input/Output

RJ-45 based LAN

Power supply input connector

Remote start

2.3.9 Portable 3000 Micro GC Onboard Battery

The 3000 Micro GC onboard battery will operate the instrument up to two hours,
depending on the application.