3000 Micro GC Operating Manual

Figure 8-1 Example bakeout method

3

Click File >> Method >> Save to save the bakeout method. When prompted,
type a unique name for the bakeout method.

4

Verify that the carrier gas is on and flowing. This will protect the column and
TCD.

5

Click Control >> Download Method to download the bakeout method.

6

Allow the method to run for the duration listed in

Table 8-1

.

7

After the bakeout is complete, load an analytical method and run a set of
calibration samples. Refer to

section 7.7 on page 7-19

and

section 7.12 on

page 7-36

.

8

Check the report. Adjust the calibration settings, retention times, and method
settings as needed.