INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 167

Advertising
background image

7 - 39

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

8

Select the Calibration check box and type 1 for Calibration level. See

Figure

7-50

.

NOTE: For the initial calibration, clear any of the boxes involving calibrations

or replicates. However, if the method was previously calibrated and the
calibrated information needs to be replaced, select Clear all
calibration
before starting.

Figure 7-50 Analysis/single level calibration window - calibration check boxes

9

Click Start after completing the window. See

Figure 7-51

.

Figure 7-51 Analysis/single level calibration window - start

When the analysis is complete, the chromatogram will be integrated, and the areas
for the peaks identified as calibration compounds will be entered into the method.
In addition, a calibration curve will be generated using these areas. The method is
now calibrated for a single level.