3000 Micro GC Operating Manual

4

Ensure that the front ferrule makes contact with the female fitting, slide the
Swagelok nut over the ferrule, then finger tighten. See

Figure 4-2

.

Figure 4-2 Vise fitting assembly

5

Push the tube fully into the female fitting.

6

Mark the Swagelok fitting with a pencil line as shown in

Figure 4-3

.

Figure 4-3 Pencil line

7

Tighten the 1/8 in. fitting 3/4 of a turn, referenced to the pencil line, using one
7/16 in. wrench to hold the fitting steady, and the other 7/16 in. wrench to turn
the fitting. See

Figure 4-4

.

Figure 4-4 Using wrenches to tighten Swagelok fittings

8

Unscrew the nut. Connect the tubing with the swaged ferrules to the pencil
marked location. Tighten the nut 1/4 turn past finger tight.

3000 Micro GC is equipped with a 1/16 in. sample inlet Swagelok connector on the
faceplate. Connecting sample inlet tubing is the same procedure as described for
carrier gas inlet tubing.

Female fitting

in vise

Front ferrule, back ferrule, and nut

Pencil line