3000 Micro GC Operating Manual

6.4 Change Carrier Gas Configuration

The type of carrier gas used for analysis can be changed through the embedded
web page.

NOTE: If a carrier gas type is changed without following this procedure, inverted

peaks will appear in the chromatogram.

1

Open the web browser.

2

In the address bar, type the current IP Address of the instrument. Press Enter.
See

Figure 6-9

.

NOTE: At the factory, the instrument IP address was set to 10.1.1.101.

Figure 6-9 Internet Explorer address bar

3

Click the Gas Type tab. A window will display that is similar to

Figure 6-10

. The

current gas configuration for installed GC modules is shown.

Figure 6-10 Example gas type tab