6.4 Change Carrier Gas Configuration
The type of carrier gas used for analysis can be changed through the embedded
web page.
NOTE: If a carrier gas type is changed without following this procedure, inverted
peaks will appear in the chromatogram.
1
Open the web browser.
2
In the address bar, type the current IP Address of the instrument. Press Enter.
See
NOTE: At the factory, the instrument IP address was set to 10.1.1.101.
Figure 6-9 Internet Explorer address bar
3
Click the Gas Type tab. A window will display that is similar to
. The
current gas configuration for installed GC modules is shown.
Figure 6-10 Example gas type tab