C
3000 Micro GC Operating Manual
11
Click Start >> Run.
12
Type ping 192.168.1.99.
This will open the Command Prompt window. The instrument should populate
the window as shown in
Figure 6-28 Fixed IP address ping
13
Type 192.168.1.99 into the web browser address bar. Press Enter. The
instrument embedded web page home page will display.
14
Click
IP Configuration.
15
Review the IP Address.
NOTE: If connection cannot be established, verify all LAN connections and
IP addresses. If there is still no communication, contact
.