7.11.2 Adding Peak and Group Information

Peaks and groups can be identified in several ways. The information can be
entered manually into the peak table, or the information can be gathered using
graphical programming tools.

7.11.2.1 Manually Add Peak Information

1

Select a channel to add peak information from the drop-down list on the tool
bar. See

Figure 7-35

.

Figure 7-35 Changing channels

2

Click Method >> Peaks/Groups.

3

The Peak/Group Tables window will display. See

Figure 7-36

. There are tabs

for Named Peaks and Groups.

Figure 7-36 Peaks/groups tables window

4

On the Named Peaks tab, enter the Name, Ret.Time, and retention time
Window for each component in the channel.

5

Repeat Step 3 for all channels. Click Analysis >> Analyze from the toolbar to
apply the changes.

6

Save the method by clicking File >> Method >> Save.