4.2.4.2 Connecting the Carrier Gas Particulate Filter

1

Remove any shipping caps on 3000 Micro GC.

2

Install one carrier gas particulate filter (PN 3150-0602) on the carrier gas
supply tubing for each carrier gas used (see

Figure 4-5

). Do not connect the

tubing to 3000 Micro GC at this time.

Figure 4-5 Carrier gas particulate filter setup

Install the carrier gas filter within 15 cm (6 in.) of the 3000 Micro GC fitting
to maximize its effectiveness.

Use a cutoff valve (not provided) as shown in

Figure 4-5

for ease of

maintenance. The valve should be "packless" (sealed to atmosphere) and
grease free. Common industrial or home-supply ball valves should not be
used.

Chemical traps on the carrier gas line may be used to protect the column.

Particulate Filter
(3150-0602)
Arrow points to
3000 Micro GC

Cutoff valve (not provided)

15 cm (6 in.)