9.1.4 Common Hardware Problems

Table 9-4 Common hardware problems

Symptom

Cause

Remedy

One carrier runs out faster
than another

External leak

Check carrier gas lines using a leak detector.
Tighten all fittings.

Internal leak

Remove the chassis cover and listen for a
hissing noise. If the noise is present, the
module is not fully seated. Use a Torx 20
screwdriver to tighten the screw connecting
the module to the baseplate. If hissing
persists, contact

INFICON

.

Power switch is blinking

Power problem

Contact

INFICON

.

3000 Micro GC will not
turn on

3000 Micro GC not
plugged in

Verify that all plugs are in (seated) correctly.

Power supply is not
supplying sufficient voltage

Test the power supply using a voltmeter.

Hardware problem

Contact

INFICON

.

Hissing sound within
instrument

Internal leak

Remove the chassis cover and listen for a
hissing noise. If the noise is present, the
module is not fully seated. Use a Torx 20
screwdriver to tighten the screw connecting
the module to the baseplate. If hissing
persists, contact

INFICON

.

Pump does not run or runs
intermittently

Sample Pump turned off

In EZ IQ, ensure the Sample Pump checkbox
is selected and set for at least 15 seconds.

Malfunctioning sample pump Listen for the noise from the pump when

taking a sample. If there is no noise, contact

INFICON

.

Standard module isn’t shown
on embedded web page

Module not configured

Commission a module through the embedded
webpage. Refer to

section 8.2.7, Commission

a New Standard GC Module, on page 8-20

.

Module had a firmware
update

If the module had a firmware upgrade, it must
be re-commissioned. Refer to

section 6.3.6,

Change GC Module Address, on page 6-8

.

Module cannot be
commissioned

C-Board Configuration
Problem

Contact

INFICON

.

DIP switch set incorrectly
(Enhanced Modules)

Ensure the DIP switch is set to the correct
address for Enhanced Modules.