Chapter 5

Accessory Installation

5.1 Sample Conditioners

3000 Micro GC can be damaged by contaminants, especially particulates and
condensing aerosols. Use an appropriate filter or sample conditioner at all times.

Table 5-1

lists the typical usage and filtration capabilities of the standard filter and

available accessories.

5.1.1 List of Sample Conditioners and IPNs

Table 5-1 List of sample conditioners, inlet pressures, and IPNs

Type

Input Pressure

to Accessory

Sample

Container/delivery

Sample Matrix

Particle

filtration

(microns)

PN

Standard external
sample inlet filter, or
direct connection

0 to 172 kPa
(0 to 25 PSI)

All

Relatively clean

and dry

10

G2801-60900

Genie Filter
Assembly

70 to 345 kPa
(10 to 50 PSI)

All

Entrained liquids

and particles

-

G2817A

Pressure Reducer

345 to 6895 kPa

(Ambient to

1000 PSI)

All

C6+ components

<0.5 mole%

-

G2815A

Pressure Reducer
and Genie Filter
Assembly

<3477 kPa
(<500 PSI)

All

Entrained liquids

and particles,

C6+ components

<0.5 mole%

G2816A