51
9-
P1
C
3000 Micro GC Operating Manual
7.7 Run a Sample
Once a method has been downloaded to 3000 Micro GC, a data acquisition run
can be performed.
1
To start the run, click Control >> Single Run. The Single Run Acquisition
window will be displayed. See
.
Figure 7-23 Single run acquisition window
2
Type a Sample ID in the first box. See
.
3
The method will default to the method that is open at the time. To change the
method, click the folder icon located on the right of the Method field. Refer to
.
4
Select a path for storage of data files by clicking the folder icon located on the
right of the Data path field. See
5
The Data file
box is used to create the filename. To automatically include
certain descriptors such as Sample ID, Method Name, and Date and Time,
click the blue arrow located to the right of the Data file
field. Select the desired
6
Type the Number of reps. See
.
7
Click Start to begin the run. Data will be displayed in the chromatogram
window. See
.