3000 Micro GC Operating Manual

Chromatograms not
separating/no data is
showing

3000 Micro GC requires
power cycle

Restart 3000 Micro GC as instructed in

section 6.3.2, Restart, on page 6-8

.

View is set to Overlay

In EZ IQ, click View

>>

Tile All Data to

generate separate chromatograms.

EZ IQ requires restart

Close EZ IQ and re-open. Restart the
computer if necessary.

Chromatograms not
separating/no data is
showing (continued)

Drivers are corrupt

Re-configure the drivers through EZ IQ
Config.

Calibration is poor

Contamination in calibration
or carrier gas

Ensure the carrier gas purity is between
99.995% and 99.9995%.
Add hydrocarbon/moisture/oxygen traps or
particulate filters to the carrier gas line to
improve purity.

Ensure there are no contaminants in the
calibration gas.

Instrument requires power
cycle

Restart 3000 Micro GC as instructed in

section 6.3.2, Restart, on page 6-8

.

External leak

Check carrier gas lines using a leak detector.
Tighten all fittings.

Poor integration

Ensure all of the peaks of interest are
identified and fully integrated.

Backflush Time set too
early (if applicable)

Increase the Backflush Time until all peaks of
interest appear on the chromatogram.

Table 9-1 Common chromatographic problems (continued)

Symptom

Possible Cause

Remedy