8.2.4 Decommission a Standard GC Module

A Standard GC module must be decommissioned within the embedded web page
before it can be removed. 3000 Micro GC internally communicates to each
installed GC module using a unique module address (see

Figure 8-3

).

For a 1-2 channel or portable 3000 Micro GC, the module addresses are
1 and 2.

For a 3-4 channel 3000 Micro GC, the module addresses are 1, 2, 3 and 4.

Decommissioning a standard GC module will disable the GC module's address.

Figure 8-3 Default GC module serial communication addresses

NOTE: For Performance Enhanced modules, skip to

section 8.2.5, Remove a GC

Module, on page 8-11

.

NOTE: Only one Standard GC Module can be decommissioned at a time.

The procedure below describes the procedure for replacing the Standard GC
module 1 (Channel A) in a 2-channel 3000 Micro GC. The process is similar for
other configurations.

1

Open the web browser and type in the 3000 Micro GC IP Address.
The embedded webpage will display.

NOTE: At the factory, the instrument IP address was set to 10.1.1.101.

2

Click

the Utilities

tab.

3

Click Change Module Configuration. See

Figure 8-4

.

Address

1

1

2

2

3

4

Front

Front

2-channel 3000

Micro GC (top down view)

4-channel 3000

Micro GC (top down view)

Module

address

Corresponding

channel

1
2
3
4

A
B
C
D