INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 195

Advertising
background image

8 - 13

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

NOTE: In Portable 3000 Micro GC, the DC-DC converter assembly blocks access

to the GC modules. Remove the screws that secure the DC-DC converter
assembly to the side of the chassis. Without disconnecting any wires, slide
the assembly off of the standoffs and gently lift it away from the GC
modules. See

Figure 8-11

.

Figure 8-11 DC-DC converter assembly (Portable 3000 Micro GC only)

6

Loosen the screw in the inlet manifold fitting at the back of the GC module using
a T-20 screwdriver. See

Figure 8-12

.

Figure 8-12 Disconnect GC module manifold fitting

DC-DC

converter

assembly

Remove screws

Manifold
fitting
screw

2-channel 3000 Micro GC

Others are similar

Loosen screw

Advertising