11.5 NGA Checkout Gas
The Natural Gas Analyzer (NGA) Checkout Gas may be used during installation of
a 3000 Micro GC configured for natural gas composition/BTU analysis to verify
chemical performance.
lists the component concentrations.
NOTE: Do not use NGA Checkout Gas as a calibration standard for specific
natural gas applications. Since natural gas composition differs from region
to region, acquire a calibration standard that meets specific requirements.
Table 11-10 NGA checkout gas calibration standard
Components
Concentration (Mole %)
nitrogen
5.2%
methane
balance
ethane
9.0%
carbon dioxide
1.5%
propane
6.0%
isobutane
3.0%
n-butane
2.0%
isopentane
0.50%
n-pentane
0.50%
n-hexane
0.10%