11.5 NGA Checkout Gas

The Natural Gas Analyzer (NGA) Checkout Gas may be used during installation of
a 3000 Micro GC configured for natural gas composition/BTU analysis to verify
chemical performance.

Table 11-10

lists the component concentrations.

NOTE: Do not use NGA Checkout Gas as a calibration standard for specific

natural gas applications. Since natural gas composition differs from region
to region, acquire a calibration standard that meets specific requirements.

Table 11-10 NGA checkout gas calibration standard

Components

Concentration (Mole %)

nitrogen

5.2%

methane

balance

ethane

9.0%

carbon dioxide

1.5%

propane

6.0%

isobutane

3.0%

n-butane

2.0%

isopentane

0.50%

n-pentane

0.50%

n-hexane

0.10%