8.2.3 Update the Instrument Firmware
Before installing a new GC Module, the 3000 Micro GC firmware must be updated.
New GC modules have the most recent firmware installed. The most recent
3000 Micro GC firmware is included in the GC module replacement kit. To update
the instrument firmware:
1
Insert the Firmware Update CD-ROM from the GC module replacement kit.
2
Browse the CD-ROM, and open the file named
readme.htm.
3
Follow the instructions in the file.
CAUTION
During a firmware update, do not turn 3000 Micro GC off
until prompted by the update program. Turning off power
during the update can make 3000 Micro GC inoperable.