7.15.3 Running a Sequence with a Valco Valve

A sequence of multiple runs with different valve positions can be scheduled by
editing the sequence as below:

1

Click Sequence >> Edit.

2

The Sequence window will display. Vial number defines the port position for a
single run. Volume defines the sample line purge time for a single run. See

Figure 7-70

.

Figure 7-70 Sequence window

Using the above sequence table as an example:

Run #1 will switch to Port #1, purge for 1 minute, then return to the conditions
specified in Sampling tab in Method >> Instrument Setup (see

Figure 7-68

)

Run #2 will switch to Port #2, purge for 1 minute, then return to the conditions
specified in Sampling tab in Method >> Instrument Setup

Run #3 will switch to Port #3, purge for 1 minute, then return to the conditions
specified in Sampling tab in Method >> Instrument Setup