2 the remote connector, C. see – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 104

Advertising
background image

5 - 30

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

5.2 The Remote Connector

For some applications, an external triggering device may be required for sampling.
This can be accomplished by connecting 3000 Micro GC to the external triggering
device via a Remote Start Cable (PN G2801-60618). This cable has a 15 pin
connector on one end. The other end can be fitted with a desired connector type.
See

Table 5-10

and

Figure 5-38

.

When using EZ IQ control with no external trigger, runs begin and end
automatically depending on Work List status (started, stopped, or paused) and
3000 Micro GC readiness. To have an external triggering device start or cancel a
run, the Digital I/O module (PN G2847A) must be installed in 3000 Micro GC. If the
Digital I/O module is not already installed, see Installing the Digital IO Module Quick
Start Guide
.

Figure 5-38 Remote start/cancel connector pin outputs

Table 5-10 Remote start/cancel connector pin outputs

Pin

Function

Wire Color

1

Provides 5 mA for remote input

Black

2

REMOTE_START input

White

3

REMOTE_START input

Red

4

Variable GND

Green

5

Provides "5 V pull up" for logic output

Orange

6

Provides "5 V pull up" for logic output

Blue

7

Contact closure output FAULT_OUT*

White/black

8

Contact closure output READY_OUT

Red/black

9

Provides 5 mA for remote input

Green/black

10

REMOTE_CANCEL input

Orange/black

11

REMOTE_CANCEL input

Blue/black

12

Variable GND

Black/white

13

GND

Red/white

14

Contact closure output FAULT_OUT*

Green/white

15

Contact closure output READY_OUT

Blue/white

*Not implemented