8.2.7 Commission a New Standard GC Module
For Standard GC modules, a new module IP address must be assigned to the
newly installed GC module. This can be accomplished using the embedded
3000 Micro GC web page.
NOTE: For a Performance Enhanced GC module, these steps can be skipped.
Proceed to
section 8.2.8, Enable the Instrument in EZ IQ, on page 8-22
1
Turn 3000 Micro GC on
2
Wait three minutes for the instrument to fully power up.
3
In the web browser address bar, type the 3000 Micro GC IP address.
4
Click the Utilities tab.
5
Click Change Module Configuration.
Figure 8-24 Change module configuration
6
Click Set. See
.
Figure 8-25 Commission a standard GC module - set