8.2.7 Commission a New Standard GC Module

For Standard GC modules, a new module IP address must be assigned to the
newly installed GC module. This can be accomplished using the embedded
3000 Micro GC web page.

NOTE: For a Performance Enhanced GC module, these steps can be skipped.

Proceed to

section 8.2.8, Enable the Instrument in EZ IQ, on page 8-22

.

1

Turn 3000 Micro GC on

2

Wait three minutes for the instrument to fully power up.

3

In the web browser address bar, type the 3000 Micro GC IP address.

4

Click the Utilities tab.

5

Click Change Module Configuration.

Figure 8-24 Change module configuration

6

Click Set. See

Figure 8-25

.

Figure 8-25 Commission a standard GC module - set