5 review calibration, N. see – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual
Page 169
7 - 41
PN
07
4-
51
9-
P1
C
3000 Micro GC Operating Manual
7.12.5 Review Calibration
After completing a calibration, a calibration curve and associated data can be
reviewed by clicking Method >> Review Calibration. Alternatively, click the
Review Calibration icon. See
.
Figure 7-53 Review calibration window
From Review Calibration,
View the calibration curve for each calibrated peak
Change and overlap calibration curve fit types
Review the equations for curve fit types, and examine the R-squared
(goodness of fit) value for each
Temporarily remove any undesired points or outliers from the calibration
curve
View response factors, replicate areas, and standard deviation values
Use the concentration calculator to calculate amounts from manually
entered areas.
Refer to the 074-538-P1(A) EZ IQ Operation Manual for a detailed description of
each function.
The method is now calibrated and is ready to use.