3000 Micro GC Operating Manual

7.12.5 Review Calibration

After completing a calibration, a calibration curve and associated data can be
reviewed by clicking Method >> Review Calibration. Alternatively, click the
Review Calibration icon. See

Figure 7-53

.

Figure 7-53 Review calibration window

From Review Calibration,

View the calibration curve for each calibrated peak

Change and overlap calibration curve fit types

Review the equations for curve fit types, and examine the R-squared
(goodness of fit) value for each

Temporarily remove any undesired points or outliers from the calibration
curve

View response factors, replicate areas, and standard deviation values

Use the concentration calculator to calculate amounts from manually
entered areas.

Refer to the 074-538-P1(A) EZ IQ Operation Manual for a detailed description of
each function.

The method is now calibrated and is ready to use.