7.12.4 Perform a Multiple Level Calibration

If a multiple level calibration is required, sample calibration standards of different
concentration levels should be run.

NOTE: Update the Method >> Peaks/Groups table to include Level 2, Level 3,

etc.

1

Repeat Steps 1 through 7 in

section 7.12.3 on page 7-37

.

2

Select the Calibrate check box and type a value in the Calibration level box
for the corresponding level. See

Figure 7-52

.

Figure 7-52 Analysis/single level calibration window - multiple level calibration

3

Click Start.