8.3.5 Running a High Pressure Gas Sample with Liquids/Particulates

1

If the Pressure Reducer and Genie Filter Assembly (PN G2816A) are not
installed, install them as instructed in

section 5.1.5, Installing the Pressure

Reducer and Genie Filter Assembly, on page 5-12

. See

Figure 8-32

.

NOTE: The Pressure Reducer and Genie Filter Assembly can connect directly

to 3000 Micro GC or to the sample inlet filter (PN G2801-60900).

NOTE: There is no set range for the flow meter. Verify some flow is present.

Figure 8-32 Pressure reducer and Genie filter assembly sample conditioner (PN G2816A)

2

Turn 3000 Micro GC.

3

Check that the carrier gas supply is sufficient to run all of the samples.

4

If using a sample vessel, prepare the sample.

5

Connect the sample vessel or gas source to the NPT sample input fitting on the
Genie Filter Assembly.

6

Open the sample vessel valve.

7

Use EZ IQ to load the desired method and perform the analysis.

8

When the analysis is complete, close the sample vessel valve and disconnect
the vessel from the instrument.

9

Repeat steps 4 to 8 to run the next sample.