4.2.5.2 Tools Required

Two 7/16 in. wrenches, for the sample inlet filter assembly

One 5/16 in. wrench, for user sample line to sample inlet filter

4.2.5.3 Sample Inlet Filter Assembly Installation

1

Shut off any sample flow to 3000 Micro GC.

2

Allow the sample inlet to cool.

3

Disconnect the sample line to the 3000 Micro GC input fitting using the 5/16 in.
wrench, while holding the sample inlet filter with a 7/16 in. wrench.

4

Inspect the sample inlet filter assembly and verify that the filter disk is in place.
If not, place a filter disk between parts A and B of the filter body and thread the
parts together until finger-tight. See

Figure 4-8

.

Figure 4-8 Exploded view of sample inlet filter assembly G2801-60900

5

Install the sample inlet filter assembly on the 3000 Micro GC inlet using the
dual-ended ferrule. Use the two 7/16 in. wrenches to tighten the filter halves
until snug.

5a

Tighten sample inlet filter assembly to 3000 Micro GC inlet an additional
1/4 turn using 7/16 in. wrench on filter Part A.

6

Connect the sample line to the male portion of filter Part B, shown in

Figure 4-8

,

while stabilizing the sample inlet filter assembly Part B using a 7/16 in. wrench.

CAUTION

Do not overtighten the sample inlet filter assembly to
3000 Micro GC. This will damage the dual-ended ferrule.

Part B

Part A

Filter disk

Dual-ended ferrule

To GC