5.1.3.2 User Supplied Parts

Fittings for vent tubing, if used

Sample input tubing and fittings

5.1.3.3 Installation Instructions for the Genie Filter Assembly

1

Turn off 3000 Micro GC and allow the sample inlet fitting to cool.

WARNING

3000 Micro GC has inlets which can be heated to 140°C.
Touching the inlets at operating temperatures can result
in injury. Extreme care should be taken to avoid touching
the heated inlets.

2

Move the 3000 Micro GC handle fully away from the front cover.

3

Remove any fittings or hardware currently installed on the 3000 Micro GC
sample fitting.

4

Place a 1/4 in. male nut, a back ferrule, and front ferrule on the outlet tube.
Use the longer outlet tube on the portable models. See

Figure 5-2

for ferrule

orientation.

5

Loosely install a 5/16 in. nut, back ferrule, and front ferrule onto the
3000 Micro GC sample inlet fitting. Do not tighten. Tubing will be installed into
this fitting later.

Figure 5-2 Installing ferrules

1/4 in. male nut

Back ferrule

Front ferrule

5/16 in. female nut

Back ferrule

Front ferrule

5/16 in. female nut