4.2.4.3 Setting the Gas Pressure

Set the carrier gas regulator as shown in

Table 4-1

.

WARNING

Never fill the Portable 3000 Micro GC internal gas
cylinder with hydrogen. A hydrogen leak inside the
instrument can present an explosion hazard.

CAUTION

Exceeding 82 PSI for the carrier gas pressure may result
in damage to 3000 Micro GC.

CAUTION

Carrier gas pressure below 78 PSI will negatively affect
3000 Micro GC operation.

4.2.4.4 Purge the Carrier Gas Line

Ensure the gas line is not connected to the instrument. Purge carrier gas supply
lines to remove air by opening the tank and letting the carrier gas flow for
five to ten seconds.

Table 4-1 Carrier gas input pressure

Carrier Gas

Required Delivery Pressure

helium*

552 +/- 14 kPa (80 +/- 2 PSI)

hydrogen*

552 +/- 14 kPa (80 +/- 2 PSI)

argon

552 +/- 14 kPa (80 +/- 2 PSI)

nitrogen

552 +/- 14 kPa (80 +/- 2 PSI)

* Required for checkout