Caution – INFICON 3000 Micro GC Gas Analyzer Operating Manual User Manual

Page 196

Advertising
background image

8 - 14

PN

07

4-

51

9-

P1

C

3000 Micro GC Operating Manual

7

Disconnect the inlet manifold fitting from the GC module input fitting using a
1/4 in. open-ended wrench. See

Figure 8-13

.

Figure 8-13 Disconnect inlet fitting

8

Disconnect the communication cables leading from the GC modules, if present,
to the communications board connectors for the GC modules. If necessary,
disconnect the fan power cables. See

Figure 8-14

.

Figure 8-14 Disconnect communication cables

9

Tilt the back of the GC module up until the mounting flange clears the alignment
pins. Slide the GC module towards the back of the chassis until it clears the
inlet manifold frame and can be angled and removed from the front. See

Figure

8-15

.

CAUTION

Be sure to lift and remove the GC module from the front
end of the GC to avoid damage to the sample pumps in
the rear. Also, be careful to avoid damaging any nearby
wires or cables.

2-channel 3000 Micro GC

Others are similar

Channel A

Channel B

Inlet

manifold

GC

module

Disconnect

inlet fitting

Back of GC
module

Remove cable

Back view of GC modules

Communications

board

Remove cable