4.3.3 Ping the Instrument
Ensure that the instrument is turned on and is fully booted up. On the computer,
click Start >> Run and type ping 10.1.1.101. Ensure data packets are being
received, as shown in
. If data packets are not being received, check
the IP settings or consult the troubleshooting table in
Connection Problems, on page 9-8
.
NOTE: At the factory, the instrument IP address was set to 10.1.1.101.
Figure 4-12 Command prompt showing successful ping
4.3.4 Carrier Gas Configuration
At the factory, 3000 Micro GC is configured to use helium carrier gas to verify
instrument quality. If the checkout gas was ordered from INFICON and it will be
used for the instrument checkout, do not change the carrier gas software setup at
this point.
If a customized checkout gas is necessary or no checkout is required, reconfigure
the carrier gas type as needed by following the instructions in