9.1.2 Common Software Problems

Table 9-2 Common software problems

Symptom

Cause

Remedy

Method grayed out

Instrument not properly
configured

Re-configure the drivers through EZ IQ
Config
.

Incorrect carrier gas

Check the carrier gases and ensure each
carrier gas is connected to the correct
3000 Micro GC inlet. See the 3000 Micro GC
webpage to verify proper gas setup. Refer to

section 6.4, Change Carrier Gas
Configuration, on page 6-9

.

EZ IQ is installed incorrectly Re-install EZ IQ. Refer to the

074-537 EZ IQ Installation Guide.

Peaks are present, but not
identified

Integration settings not
optimized

Ensure all of the peaks of interest are
identified and fully integrated.

Peak is outside RT window

In EZ IQ, verify that the peak window contains
the desired peak.

TCD and columns turn off

Low carrier gas pressure

Ensure the carrier gas is supplying at least
78 PSI and no more that 82 PSI.

CAUTION

Exceeding 82 PSI for the
carrier gas pressure may
damage 3000 Micro GC.