8.4.3 Replacing the 7 Micron Filter in the Heated Regulator

(PN G2818A-X/G2845A/G2857A)

The 7 micron filter assembly uses a replaceable cartridge filter (PN 3150-0786).
See

Figure 8-38

.

Figure 8-38 7 Micron filter assembly

1

Turn off any sample flow to 3000 Micro GC.

2

Turn the relief valve to Vent on the Heated Regulator and purge all residual
pressure remaining in the sample line.

3

Unplug the heated sample conditioner and allow it to cool.

4

Loosen the screws that secure the Heated Regulator cover in place and
remove the cover.

5

Remove the 1/16 in. line at the bottom of the filter assembly See

Figure 8-39

.

6

Remove the nut at the bottom of the filter.

Figure 8-39 Replacing the 7 micron filter

Remove from heater
block

7 micron filter

Remove nut
Remove 1/16 in. line