3000 Micro GC Operating Manual

Only clean gases or vapors should be introduced to 3000 Micro GC. Avoid
aerosols, condensable vapors, liquids and solid particles. Install an appropriate
sample filter or conditioner. An external 10 micron filter (PN G2801-60900) is
shipped with every 3000 Micro GC. This filter must be used unless it is replaced by
another filter or sample conditioner. Contact

INFICON

for details on available

accessories.

3000 Micro GC can be controlled by a personal computer (computer) through a
standard RJ-45 connection, either directly from a computer using a cross-over
cable connection or through a Local Area Network (LAN) connection. EZ IQ
software controls all experimental settings, data collection, and data analysis.

Table 1-1 Summary of connections

Connection

Notes

Input fitting(s)

1/16 in. Swagelok

Input pressure range
Recommended range

0 to 172 kPa (0 to 25 PSI)
35 to 69 kPa (5 to 10 PSI)

Sample filtration

External 10 micron particle trap standard (not used with
accessories G2816A, G2817A, G2818A, G2819A-X

1

,

G2845A-X

1

, or G2846A-X

1

, G2857A-X

1

, G2858A-X

1

)

1

Where X is the country specific power cord number:

1: China
2: Europe
3: US
4: Japan
5: UK/HK/SG/MY
6: Australia/NZ
7: Korea

Instrument control

EZ IQ software

Carrier gas inlet fitting

1/8 in. Swagelok

Vent gas fitting

1/8 in. Luer-Lok

®