8.4 Consumable Replacement Procedures

8.4.1 Replacing the External 10 Micron Sample Inlet Filter Disk

1

Shut off any sample flow to the GC.

2

Wait for the 3000 Micro GC inlet to cool.

3

Disconnect any sample line or conditioner to the sample inlet filter.

4

Using a 5/16 in. wrench, remove Part B. See

Figure 8-36

.

Figure 8-36 Replacing the sample inlet filter disk

5

Remove the used filter disk from Part A.

6

Replace the used filter disk with a new one (PN 5183-4652).

7

Reassemble the sample inlet filter. Turn Part B clockwise until finger-tight, then
use a wrench to tighten an additional 1/4 turn.

CAUTION

Do not overtighten the sample inlet filter assembly to
3000 Micro GC. This will damage the dual-ended ferrule.

8

Attach sample line or conditioner to sample inlet filter.

9

Resume sample flow as needed.

Part B

Part A

Dual-end ferrule

Filter disk

To GC