Figure 5-32 Installing the heated sample instroduction accessory on 3000 Micro GC
3
Tighten the 3000 Micro GC cover screws.
4
Insert the transfer line into the 3000 Micro GC sample inlet fitting as far as
possible, then back it out about 1 mm.
5
Finger tighten the female nut, then tighten an additional 1/8 turn with the
5/16 in. open-ended wrench.
6
If the transfer line is kinked, remove the sample introduction accessory and trim
about 1 mm off the end of the transfer line. Make the cut square and
perpendicular. Reassemble. See
.
Figure 5-33 Trimming the transfer line
7
Install and tighten the upper left cover mounting screw using T20 Torx
screwdriver.
8
Tighten the remaining cover screws with a T20 Torx screwdriver.
9
Wrap one of the two pieces of insulation around the transfer line. Retain the
remaining transfer line, insulation and extra mounting hardware as spares.
Sample inlet fitting
Transfer line
Trim 1 mm of
transfer line
Transfer line