3000 Micro GC Operating Manual
2
The Analysis/Single Level Calibration window will display.
Figure 7-48 Analysis/single level calibration window - analysis information
3
Type the sample identification in the Sample ID box, if not already present. See
.
4
Select the name of the method to be calibrated, if not already present, by
clicking the Open file icon adjacent to the box. See
.
5
Select the data path name, if not already present, by clicking the
Open file
button adjacent to the field. See
6
Select the name of the calibration data file in the Data file field. Select the file
name, if not already present, by clicking the Open file button adjacent to the
field. See
.
7
Leave the
Amount values set to 1. A value of 1 corresponds to a single level
calibration. See
. For more information on each parameter within
the Amount values frame, refer to the 074-538 EZ IQ Operation Manual.
Figure 7-49 Analysis/single level calibration window - amount values