5.1.4.3 Installation Instructions for the Pressure Reducer

1

Turn off 3000 Micro GC and allow the sample inlet fitting to cool.

WARNING

3000 Micro GC has inlets which can be heated to 140°C.
Touching the inlets at operating temperatures can result
in injury. Extreme care should be taken to avoid touching
the heated inlets.

2

Move the 3000 Micro GC handle fully away from the front cover.

3

Remove any fittings or hardware currently installed on the 3000 Micro GC
sample fitting.

NOTE: The external 10 micron sample inlet filter may be left in place. Refer to

section 5.1.2 on page 5-3

.

4

Loosely install a 5/16 in. female nut, back ferrule, and front ferrule onto the
sample inlet fitting or the 10 micron sample inlet filter. See

Figure 5-9

. Do not

tighten. The outlet tube will be installed into this fitting later.

Figure 5-9 Installing ferrules

5

Insert the outlet tube into the top 1/16 in. fitting on the flow controller (see

Figure

5-10

). Tighten 1/2 turn past finger tight using the 1/4 in. open-ended wrench,

then loosen 1/4 turn, so that the tube is loose but will not disengage.

Figure 5-10 Installing the sample outlet tube

Back ferrule

Front ferrule

5/16 in. female nut

Sample in

Vent out

Install outlet
tube here