5.1.6.3.4 Installing the Transfer Line

1

The sample outlet fitting on the heated sample introduction accessory contains
a nut and ferrule set and should be ready for transfer line installation. See

Figure 5-27

.

Figure 5-27 Sample outlet fitting nut

2

Before mounting the heated sample introduction accessory, the transfer line
must be inserted into the sample outlet fitting. Verify that the transfer line can
slide into the nut and ferrule set. Loosen the nut if necessary.

3

Insert the transfer line into the sample outlet fitting as far as possible, then back
it out about 1 mm.

4

Finger-tighten the nut, then make an additional 1/8 turn with the 5/16 in.
open-ended wrench. See

Figure 5-28

.

Figure 5-28 Transfer line installed on a heated regulator

Quick disconnect fitting

Sample outlet fitting

Back view of heated sample conditioners
(bracket for 2-channel 3000 Micro GC)