Feedback device – Yaskawa iQpump Programming Manual User Manual

Page 14

Advertising
background image

14

YASKAWA

TM.iQp.02 iQpump Drive Programming Manual

Feedback Device

The iQpump Controller requires a feedback device (e.g. Pressure transducer, flow meter, etc.) to perform automatic system regulation.

Any analog 0~10 V or 4-20 mA feedback device can be used in combination with the iQpump controller.

Connecting Your Feedback Device to the iQpump Controller

Note: The factory default setting for the iQpump controller is 4~20 mA feedback device connected to

analog input A2.

To successfully operate the iQpump drive remotely, an external run command must be received by the

Drive. Parameter b1-02 specifies from where the run command will be accepted.

Although the Run Source and the Reference Source (b1-01) are normally taken from the same source (e.g.

digital operator, terminals or serial communication), this is not always the case.

To issue a run command from the digital operator: Set b1-02 = “0: Operator,” and use the HAND and

OFF buttons to start and stop the Drive.

To issue the run command from the terminals: Set b1-02 = “1: Terminals,” and select between 2-wire

and 3-wire control operation by doing the following:

2-Wire Control The factory default setting is for 2-wire operation. In the 2-wire configuration a closure between S1 and SN will be

interpreted as a Forward Run command by the Drive.

Figure 1.7

Figure 7 2-Wire Control

3-Wire Control When any of the multi-function digital input parameters, H1-01 through H1-05, is set to 0, terminals S1 and S2 become

Run and Stop, respectively. The multi-function digital input that was set to 0 will function as a Forward/Reverse input for the iQpump

Drive. When the Forward/Reverse input is open the iQpump drive will run in the Forward direction and when the input is closed, the

iQpump drive will run in the Reverse direction.

In 3-wire operation a momentary closure (> 50mS) of S1 will cause the iQpump drive to run provided that S2 is held closed. The iQpump

drive will stop anytime the S2-SN connection is broken. If the 3-wire configuration is implemented via a 3-wire Initialization (A1-03 =

“3330: 3-Wire Initial”), then terminal S3 becomes the Forward/Reverse input.

Note: Reverse operation is disabled in the iQpump drive; however, in 3-wire control, one of the multi-function digital inputs

needs to be programmed to 0. Otherwise, the 3-wire control will not work.

Figure 1.8

Figure 8 3-Wire Control

To issue a run command via serial communication: Set b1-02 = “2: Serial Com” and connect the RS-485/422 serial communication

cable to R+, R-, S+, and S- on the removable terminal block.

• Modbus Plus Option Card CM071

Manual: IG.AFD.17

• Modbus TCP/IP Option Card CM090

Manual: IG.AFD.25

• EtherNet/IP Option Card CM092

Manual: IG.AFD.26

S1

S2

SN

FWD Run/Stop

REV Run/Stop

3-wire control

Stop switch
(NC contact)

Operation switch
(NO contact)

Run command
(run on momentary close)
Stop command
(stop on momentary open)

Sequence input common

SN

Advertising