RKI Instruments EAGLE 2 Manual User Manual

Page 102

Advertising
background image

92 • Calibrating Using the Auto Calibration Method

EAGLE 2 Operator’s Manual

12. Press and release the POWER ENTER RESET button. The following

screen appears.

13. If you select YES by pressing and releasing the AIR ▲ YES button, the

changes that you made will be saved in the EAGLE 2’s memory as the
new auto calibration gas values.

If you select NO by pressing and releasing the DISPLAY ADJUST NO
button, the changes you made will be used for any calibrations performed
during the current operating session only. The EAGLE 2 will delete the
changes when the unit is turned off and will load the previous set of auto
calibration values when it is turned on again.

14. When you make your selection and press the desired button, the unit

returns to the Calibration Gas Values Screen.

15. Press and release the POWER ENTER RESET button to proceed to the

Calibration In Process Screen. CAL IN PROCESS is flashing.

If you do not want to proceed with the calibration, press and release the
DISPLAY ADJUST NO button to return to the Calibration Gas Values
Screen.

If you do want to continue with the calibration, proceed to the next step.

16. Connect the gas bag tubing to the regulator’s or dispensing valve’s hose

barb fitting.

DO YOU WANT TO

STORE NEW VALUE(S)

IN MEMORY FOR

FUTURE CALIBRATIONS?

PRESS YES OR NO

CAL GAS VALUES
CH4 50 %LEL
OXY 12.0 vol%
H2S 25.0 ppm

CO 50 ppm

ENTER TO BEGIN CAL

CAL IN PROCESS

CH4 0 %LEL

OXY 20.9 vol%

H2S 0.0 ppm

CO 0 ppm

ENTER WHEN DONE

Advertising