RKI Instruments EAGLE 2 Manual User Manual

Page 282

Advertising
background image

272 • Start Up, Bar Hole Mode

EAGLE 2 Operator’s Manual

NOTE:

If no button is pressed for 20 seconds, the unit will proceed
into whichever mode has the cursor in front of it.

9. The Active Gases Screen appears for a few seconds indicating which

channels are active and their target gas.

10. The gas alarm setpoints are displayed by three screens in sequence: the

Low Alarm Screen, High Alarm Screen, and STEL/TWA Alarm Screen.
Each screen remains on the LCD for three seconds.

11. After the alarm screens, if CAL REMINDER is turned on, the screen that

appears next depends on how CAL PAST DUE ACT is set in the Setup
Mode Menu (see “Updating the Calibration Past Due Action Setting” on
page 123).

ACTIVE GASES

CH4 OXY
H2S CO

CH4

A CH4 10 %LEL

L L OXY 19.5 vol%

O A H2S 10.0 ppm

WR CO 25 ppm

M

S

CH4 10 vol%

A CH4 50 %LEL

HL OXY 23.5 vol%

I A H2S 30.0 ppm

GR CO 50 ppm

HM

S

CH4 50 vol%

ALARMS STEL & TWA

H2S(ppm) 15.0 10.0

CO (ppm) 200 25

CH4(vol%) OFF OFF